This guide covers the field calibration of acoustic sensors for Soft dB monitoring stations. Unlike laboratory calibration, field calibration is a simple, quick procedure that should be performed periodically to check and verify the sensitivity of the acoustic sensor against the environmental conditions present in the field.
Field calibration ensures the reliability of the measured acoustic levels. It is not a substitute for the more detailed laboratory calibration, which should ideally be performed annually, or in accordance with local regulations or specific project requirements.
The measurement stations covered in this guide are:
You can tell the two station types apart by the identification label on the environmental housing. Note that the shape of the housing and the acoustic sensor — including its environmental protection — may vary from one version to another.
Two calibration methods are described:
The equipment needed to perform a calibration depends on the type of monitoring station.
Before calibrating the acoustic sensor of a Soft dB monitoring station, make sure you have all the necessary equipment on hand, and pay particular attention to the instrument's components. The calibration procedures below refer to these specific components in their instructions.
The figure below shows the different components of an acoustic sensor and its mounting. Note that some microphones have environmental protections that may differ from the images shown in this guide.

Identification information is located on the acoustic sensor's mounting bracket. This information is also available on the microphone's environmental-protection panel.

It is strongly recommended to use the same reference source (calibrator) across calibrations. If a different source is used, be sure to note the serial number of the reference source used during each calibration.
The reference source (calibrator) displays useful information such as product information, frequency, reference levels, and the ON indicator.

This procedure is provided as a general reference tool. Always ensure that you comply with the current standards and regulations applicable to the specific requirements of your project.
Calibration lets you verify the quality of the levels measured between two calibrations. Measurement quality can be compromised by a significant deviation in microphone sensitivity, so it is essential to confirm that your measurements remain accurate.
For optimal conditions, perform monthly calibrations when certain weather conditions are met:
It is the client's responsibility to complete the calibration register during each calibration, particularly for manual-mode calibrations. This log is generally required when the data measured by the stations needs to be analyzed, and it serves as a record of the measured noise levels and sensitivities. For data accuracy, calibrate the acoustic sensors at least once a month, and avoid handling the measuring instruments on rainy or windy days — weathering of the microphone can affect the quality of the calibrations and the accuracy of the measurements.
Because of the equipment's general operating range, calibrations in temperatures below −10 °C are not recommended. If calibration below −10 °C is required, first confirm that the calibrator you are using is approved for operation at those temperatures.
This is the method Soft dB primarily recommends. The wizard guides you through each step, and when the monitoring station is online the calibration data is written to a database automatically — so you don't have to record the parameters manually. The procedure differs slightly between the two station families; follow the one that matches your station.
| Step | Watch™ Series | Pro Station™ Series |
|---|---|---|
| Connect a tablet or computer to the station and open the Watch™ Utility (Watch™ Series) or Mezzo Noise Monitor (Pro Station™ Series) software. | Watch™, step 1 | Pro Station™, step 1 |
| Record in the calibration register the calibrator's model and serial number. Only required when a different calibrator is used for each calibration. | See the calibration register | |
| Start the calibration wizard and follow the on-screen instructions in order. | Watch™ procedure | Pro Station™ procedure |
With the touchscreen connected to the station, open the Mezzo Noise Monitor software on the measuring station. In the Setup tab, click the Calib Wizard button to open the Calibration Wizard.

Verify that the Sound level and Frequency parameters are set to the levels indicated on the calibrator, and that the calibrator is activated.

Once complete, click the Next button.
Follow the on-screen instructions: remove the windscreen (foam ball) from the microphone, install the calibrator, and start the calibrator by pressing the ON button.

Once complete, click the Next button.
A spike at the calibration frequency (here, 1000 Hz) should be visible, indicating that the calibrator is working. Wait a few moments for the calibration to be carried out.

The calibration has completed successfully when all indicators show PASS.

If a result reads FAIL, review the parameter definitions in Understanding a FAIL result definition table below.
| Parameter | Unit | Definition |
|---|---|---|
| Reference Sensitivity | mV/Pa | Microphone reference sensitivity as shown on the microphone calibration certificate. This is the level considered to be the reference value. |
| As-Found Sensitivity (sensitivity before calibration) |
mV/Pa | Microphone sensitivity recorded in the sound level meter prior to field calibration. It represents the sensitivity setting used by the instrument at the time the check is performed. |
| As-Found SPL (level measured before calibration) |
dB SPL | The sound pressure level displayed by the sound level meter when the acoustic calibrator is installed on the microphone, before any adjustment. This lets the initial reading be compared to the reference level produced by the calibrator before proceeding with the calibration. |
| Calibrated Sensitivity (sensitivity after calibration) |
mV/Pa | New sensitivity value calculated and stored in the sound level meter so that the measured level matches the reference level of the acoustic calibrator. This value is then used for field measurements. The calibrated SPL is not shown, since it will be equivalent to the reference SPL after calibration. |
If the As-Found parameters fail, it may indicate a problem with the previous calibration or with the microphone's communication chain (for example, a cut cable, physical contamination, or a calibrator that is switched off).
If Calibrated Sensitivity fails, check that the calibrator and microphone are working properly (cables, connectors, and so on). If no apparent problem is found, contact Soft dB. The acoustic sensor and/or calibrator may need to be returned to Soft dB for inspection, potential repair, and/or recertification, if applicable.
If the calibration is successful, click the Next button.
Follow the on-screen instructions to remove the calibrator and reinstall the windscreen (foam ball).

To complete the process:
Using a Windows computer or tablet, open the Watch™ Utility software and connect the computer to the measuring station.

Click the Calibration Wizard button in the SLM - Microphone column to open the Calibration Wizard.

Verify that the Sound level and Frequency parameters are set to the levels indicated on the calibrator, and that the calibrator is activated.

Once complete, click the Next button.
Follow the on-screen instructions: remove the windscreen (foam ball) from the microphone, install the calibrator, and start the calibrator by pressing the ON button.

Once complete, click the Next button.
A spike at the calibration frequency (here, 1000 Hz) should be visible, indicating that the calibrator is working. Wait a few moments for the calibration to be carried out.

The calibration has completed successfully when all indicators read PASS.

If a result reads FAIL, review the parameter definitions in Understanding a FAIL result definition table below.
| Parameter | Unit | Definition |
|---|---|---|
| Reference Sensitivity | mV/Pa | Microphone reference sensitivity as shown on the microphone calibration certificate. This is the level considered to be the reference value. |
| As-Found Sensitivity (sensitivity before calibration) |
mV/Pa | Microphone sensitivity recorded in the sound level meter prior to field calibration. It represents the sensitivity setting used by the instrument at the time the check is performed. |
| As-Found SPL (level measured before calibration) |
dB SPL | The sound pressure level displayed by the sound level meter when the acoustic calibrator is installed on the microphone, before any adjustment. This lets the initial reading be compared to the reference level produced by the calibrator before proceeding with the calibration. |
| Calibrated Sensitivity (sensitivity after calibration) |
mV/Pa | New sensitivity value calculated and stored in the sound level meter so that the measured level matches the reference level of the acoustic calibrator. This value is then used for field measurements. The calibrated SPL is not shown, since it will be equivalent to the reference SPL after calibration. |
If the As-Found parameters fail, it may indicate a problem with the previous calibration or with the microphone's communication chain (for example, a cut cable, physical contamination, or a calibrator that is switched off).
If Calibrated Sensitivity fails, check that the calibrator and microphone are working properly (cables, connectors, and so on). If no apparent problem is found, contact Soft dB. The acoustic sensor and/or calibrator may need to be returned to Soft dB for inspection, potential repair, and/or recertification, if applicable.
If the calibration is successful, click the Next button.
Follow the on-screen instructions to remove the calibrator and reinstall the windscreen (foam ball).

To complete the process:
Manual calibration is the alternative to the wizard and is documented separately for each station type. Throughout the procedure, record the data you obtain in the calibration register at the end of this guide. The flow is the same for both station families: shared preparation, then the station-specific procedure (Pro or Watch), then shared finalization.
| Step | Watch™ Series | Pro Station™ Series |
|---|---|---|
| Stop measurement. | Preparation, step 1 | |
| Record in the calibration register: calibrator's model and serial number, sound level meter's serial number, nominal sensitivity written on the microphone, and calibrator's reference level. | See the register, and preparation, step 2 | |
| Install and start the calibrator. | Preparation, steps 3–5 | |
| Note the microphone sensitivity registered in the software. | Watch™, step 2 | Pro Station™, steps 1–3 |
| Check the measured level and record it in the logbook.* Compare the measured level to the calibrator's reference level.* | Watch™, step 3 | Pro Station™, step 4 |
| Restore nominal microphone sensitivity. | Watch™, step 4 | Pro Station™, step 5 |
| Remove the calibrator; note the measured level in the register and compare it to the calibrator's reference level. | Watch™, step 7 | Pro Station™, step 6 |
| Calibrate the microphone. | Watch™, step 7 | Pro Station™, step 6 |
| Note the new sensitivity in the calibration register. | See the register | |
| Finalize the calibration and resume measurement. | Finalization procedure | |
* These steps are only required during periodic calibration.
These steps are common to both station types. Complete them first, then continue with the procedure for your station.
Using the monitoring software, stop measurements.
Watch™ Series

Pro Station™ Series

Verify that the date of the last laboratory calibration — located at the base of the microphone (see Figure B) — is less than one year old. Perform the same check for the reference source.
Remove the windscreen (foam ball) at the microphone base by carefully unscrewing it counter-clockwise. Be careful not to squeeze the windscreen during removal.
Important. Soft dB cannot be held responsible for damage caused by improper handling of instruments during calibration.

Install the reference source at the end of the microphone until the microphone locks in place. Push the microphone into the receiver base of the reference source and rotate slightly clockwise.

Wait 90 seconds after installing the reference source, then start the device by pressing the On/Off button (this may vary by model; see the example in the red box).
Now continue with the procedure for your station: Pro Station™ Series: Step 6 or Watch™ Series: Step 7. Then complete the shared finalization.

The following steps make up Step 6 of the manual procedure. Perform them for Pro Station™ Series stations only.
With the touchscreen connected to the station, open the Mezzo Noise Monitor software on the measuring station and click the Stop button to stop the measurement.
Select the Setup tab. In the Setup tab, click on the Sound Input button or OPL Input button to open the interface for the type of microphone installed.

Check that the sound level generated by the reference source is lower than the saturation (overload) level of the acoustic sensor displayed in the Peak Overload field (Regular microphone) and Dynamic Range field (SAS microphone).


Confirm the microphone information displayed in the interface.


Verify that the microphone's serial number, manufacturer name, and model number are correct. This information is on the label at the base of the microphone.
Note the sensitivity value displayed in the sensitivity field in the calibration register (before calibration) and enter it in the Sensitivity — Before column.
Click the Calibrate button to open the calibration and level-check interface.

). It is strongly recommended to verify the reference level against the technical specifications of the reference source and the microphone.

Level check:

If the data is within tolerances, proceed to the next step.
Calibrate:

If the data is within tolerances, click OK to save this calibration.
Continue to the finalization steps.
The following steps make up Step 7 of the manual procedure. Perform them for Watch™ Series stations only.
Using a Windows computer or tablet, open the Watch™ Monitoring Utility software and connect the computer to the measurement station.

Click the Sensor Setup button in the SLM - Microphone column of the Watch™ Utility screen to open the Sensor Setup menu.

Enter the microphone sensitivity value displayed in the Sensitivity field in the calibration register.

Check that the Dynamic Range value displayed in the Dynamic Value field is higher than the reference source level. Click the Calibrate button to open the calibration interface.

As-found level check and restore factory calibration:

Post-restore level check:

Calibrate:

Continue to finalization steps.
Once the fields in the calibration register are filled in, finalize the calibration of the acoustic sensor.
Remove the reference source (calibrator) by rotating it slightly clockwise.
Important. Be careful not to unscrew the microphone membrane.

Reinstall the windscreen (foam ball) by rotating it clockwise onto the membrane.
Important. Make sure the windscreen is dry and fully covers the microphone.

Using the monitoring software, restart measurements.
Watch™ Series

Pro Station™ Series

The calibration register is your record of each calibration: the equipment used, the levels measured before and after, and the resulting sensitivities. Complete it during every calibration — especially for manual-mode calibrations — as described in Recommendations & best practices.

Calibration register — the form to complete during each calibration.
Register legend — colour-coded fields (for example, the green box for the sensitivity before calibration and the orange box for the measured level) correspond to the highlighted areas referenced in the procedures above.