Field Calibration Guide for Watch™ and Pro Station™ Series

Field Calibration Guide for Watch™ and Pro Station™ Series

Calibration Guideline Scope

This guide covers the field calibration of acoustic sensors for Soft dB monitoring stations. Unlike laboratory calibration, field calibration is a simple, quick procedure that should be performed periodically to check and verify the sensitivity of the acoustic sensor against the environmental conditions present in the field.

Field calibration ensures the reliability of the measured acoustic levels. It is not a substitute for the more detailed laboratory calibration, which should ideally be performed annually, or in accordance with local regulations or specific project requirements.

The measurement stations covered in this guide are:

  • Watch™ Series
  • Pro Station™ Series

You can tell the two station types apart by the identification label on the environmental housing. Note that the shape of the housing and the acoustic sensor — including its environmental protection — may vary from one version to another.

Two calibration methods are described:

  • Automatic calibration, using the calibration wizard. This is the recommended method. It guides you through a simple step-by-step procedure and, when the monitoring station is online, calibration data is recorded to a database automatically, so there's no need to log parameters by hand.
  • Manual calibration, described separately for each station type, as an alternative method.

Before You Begin

Required Equipment

The equipment needed to perform a calibration depends on the type of monitoring station.

Watch™ Series

  • Reference source or calibrator (1000 Hz, 94 dB — exact value to be validated against the reference source)
  • Computer with a micro-USB cable
  • Latest version of the Watch™ Utility software installed on the computer: softdb.com/software.php?watch

Pro Station™ Series

  • Reference source (1000 Hz, 94 dB or 114 dB — exact value to be validated against the reference source)
  • Compatible USB touchscreen

Acoustic Sensor Components

Before calibrating the acoustic sensor of a Soft dB monitoring station, make sure you have all the necessary equipment on hand, and pay particular attention to the instrument's components. The calibration procedures below refer to these specific components in their instructions.

The figure below shows the different components of an acoustic sensor and its mounting. Note that some microphones have environmental protections that may differ from the images shown in this guide.

Figure A — Components of the acoustic sensor and its mounting.

Acoustic Sensor Identification

Identification information is located on the acoustic sensor's mounting bracket. This information is also available on the microphone's environmental-protection panel.

Figure B — Where to find the acoustic sensor identification information.

Reference Source Description

It is strongly recommended to use the same reference source (calibrator) across calibrations. If a different source is used, be sure to note the serial number of the reference source used during each calibration.

The reference source (calibrator) displays useful information such as product information, frequency, reference levels, and the ON indicator.

Figure C — The reference source (calibrator) and its display.

Recommendations & Best Practices

This procedure is provided as a general reference tool. Always ensure that you comply with the current standards and regulations applicable to the specific requirements of your project.

Calibration lets you verify the quality of the levels measured between two calibrations. Measurement quality can be compromised by a significant deviation in microphone sensitivity, so it is essential to confirm that your measurements remain accurate.

For optimal conditions, perform monthly calibrations when certain weather conditions are met:

  • Wind speed is less than 20 km/h;
  • Humidity is below 90%;
  • Dry conditions, with no precipitation;
  • The ambient temperature stays within the tolerance limits specified by the manufacturer of the measuring equipment (acoustic sensor and reference source);
  • The ambient noise around the microphone is stable. Noise that is too high will affect the calibration and can skew the results. If the ambient noise is high, perform the acoustic calibration at a higher level (for example, 114 dBA), provided the microphone is capable of measuring such levels.

It is the client's responsibility to complete the calibration register during each calibration, particularly for manual-mode calibrations. This log is generally required when the data measured by the stations needs to be analyzed, and it serves as a record of the measured noise levels and sensitivities. For data accuracy, calibrate the acoustic sensors at least once a month, and avoid handling the measuring instruments on rainy or windy days — weathering of the microphone can affect the quality of the calibrations and the accuracy of the measurements.

Because of the equipment's general operating range, calibrations in temperatures below −10 °C are not recommended. If calibration below −10 °C is required, first confirm that the calibrator you are using is approved for operation at those temperatures.

Automatic Calibration Using Calibration Wizard

This is the method Soft dB primarily recommends. The wizard guides you through each step, and when the monitoring station is online the calibration data is written to a database automatically — so you don't have to record the parameters manually. The procedure differs slightly between the two station families; follow the one that matches your station.

Summary of Steps

Step Watch™ Series Pro Station™ Series
Connect a tablet or computer to the station and open the Watch™ Utility (Watch™ Series) or Mezzo Noise Monitor (Pro Station™ Series) software. Watch™, step 1 Pro Station™, step 1
Record in the calibration register the calibrator's model and serial number. Only required when a different calibrator is used for each calibration. See the calibration register
Start the calibration wizard and follow the on-screen instructions in order. Watch™ procedure Pro Station™ procedure

Automatic Calibration - Pro Station™ Series

  1. With the touchscreen connected to the station, open the Mezzo Noise Monitor software on the measuring station. In the Setup tab, click the Calib Wizard button to open the Calibration Wizard.

    Mezzo Noise Monitor Setup tab with the Calibration Wizard icon highlighted in a red box.
  2. Verify that the Sound level and Frequency parameters are set to the levels indicated on the calibrator, and that the calibrator is activated.

    Calibration Wizard parameter screen showing the sound level and frequency fields set to the calibrator's values.

    Once complete, click the Next button.

  3. Follow the on-screen instructions: remove the windscreen (foam ball) from the microphone, install the calibrator, and start the calibrator by pressing the ON button.

    On-screen instruction to remove the windscreen, install the calibrator, and switch it on.

    Once complete, click the Next button.

  4. A spike at the calibration frequency (here, 1000 Hz) should be visible, indicating that the calibrator is working. Wait a few moments for the calibration to be carried out.

    Frequency spectrum showing a clear peak at 1000 Hz during calibration.
  5. The calibration has completed successfully when all indicators show PASS.

    Calibration results screen with all indicators reading PASS.

    If a result reads FAIL, review the parameter definitions in Understanding a FAIL result definition table below.

    Understanding a FAIL result

    Parameter Unit Definition
    Reference Sensitivity mV/Pa Microphone reference sensitivity as shown on the microphone calibration certificate. This is the level considered to be the reference value.
    As-Found Sensitivity
    (sensitivity before calibration)
    mV/Pa Microphone sensitivity recorded in the sound level meter prior to field calibration. It represents the sensitivity setting used by the instrument at the time the check is performed.
    As-Found SPL
    (level measured before calibration)
    dB SPL The sound pressure level displayed by the sound level meter when the acoustic calibrator is installed on the microphone, before any adjustment. This lets the initial reading be compared to the reference level produced by the calibrator before proceeding with the calibration.
    Calibrated Sensitivity
    (sensitivity after calibration)
    mV/Pa New sensitivity value calculated and stored in the sound level meter so that the measured level matches the reference level of the acoustic calibrator. This value is then used for field measurements. The calibrated SPL is not shown, since it will be equivalent to the reference SPL after calibration.

    If the As-Found parameters fail, it may indicate a problem with the previous calibration or with the microphone's communication chain (for example, a cut cable, physical contamination, or a calibrator that is switched off).

    If Calibrated Sensitivity fails, check that the calibrator and microphone are working properly (cables, connectors, and so on). If no apparent problem is found, contact Soft dB. The acoustic sensor and/or calibrator may need to be returned to Soft dB for inspection, potential repair, and/or recertification, if applicable.

    If the calibration is successful, click the Next button.

  6. Follow the on-screen instructions to remove the calibrator and reinstall the windscreen (foam ball).

    On-screen instruction to remove the calibrator and reinstall the windscreen.

    To complete the process:

    • Click the Finish and Start button so that measurement resumes automatically.
    • Click the Finish without Starting button so that the calibration process can be completed without measurement resuming automatically.
    • Click the Cancel button to cancel the calibration process (for example, after a handling error).

Automatic Calibration - Watch™ Series

  1. Using a Windows computer or tablet, open the Watch™ Utility software and connect the computer to the measuring station.

  2. Click the Calibration Wizard button in the SLM - Microphone column to open the Calibration Wizard.

  3. Verify that the Sound level and Frequency parameters are set to the levels indicated on the calibrator, and that the calibrator is activated.

    Once complete, click the Next button.

  4. Follow the on-screen instructions: remove the windscreen (foam ball) from the microphone, install the calibrator, and start the calibrator by pressing the ON button.

    On-screen instruction to remove the windscreen, install the calibrator, and switch it on.

    Once complete, click the Next button.

  5. A spike at the calibration frequency (here, 1000 Hz) should be visible, indicating that the calibrator is working. Wait a few moments for the calibration to be carried out.

    Frequency spectrum showing a clear peak at 1000 Hz during calibration.
  6. The calibration has completed successfully when all indicators read PASS.

    Calibration results screen with all indicators reading PASS.

    If a result reads FAIL, review the parameter definitions in Understanding a FAIL result definition table below.

    Understanding a FAIL result

    Parameter Unit Definition
    Reference Sensitivity mV/Pa Microphone reference sensitivity as shown on the microphone calibration certificate. This is the level considered to be the reference value.
    As-Found Sensitivity
    (sensitivity before calibration)
    mV/Pa Microphone sensitivity recorded in the sound level meter prior to field calibration. It represents the sensitivity setting used by the instrument at the time the check is performed.
    As-Found SPL
    (level measured before calibration)
    dB SPL The sound pressure level displayed by the sound level meter when the acoustic calibrator is installed on the microphone, before any adjustment. This lets the initial reading be compared to the reference level produced by the calibrator before proceeding with the calibration.
    Calibrated Sensitivity
    (sensitivity after calibration)
    mV/Pa New sensitivity value calculated and stored in the sound level meter so that the measured level matches the reference level of the acoustic calibrator. This value is then used for field measurements. The calibrated SPL is not shown, since it will be equivalent to the reference SPL after calibration.

    If the As-Found parameters fail, it may indicate a problem with the previous calibration or with the microphone's communication chain (for example, a cut cable, physical contamination, or a calibrator that is switched off).

    If Calibrated Sensitivity fails, check that the calibrator and microphone are working properly (cables, connectors, and so on). If no apparent problem is found, contact Soft dB. The acoustic sensor and/or calibrator may need to be returned to Soft dB for inspection, potential repair, and/or recertification, if applicable.

    If the calibration is successful, click the Next button.

  7. Follow the on-screen instructions to remove the calibrator and reinstall the windscreen (foam ball).

    On-screen instruction to remove the calibrator and reinstall the windscreen.

    To complete the process:

    • Click the Finish and Start button so that measurement resumes automatically.
    • Click the Finish without Starting button so that the calibration process can be completed without measurement resuming automatically.
    • Click the Cancel button to cancel the calibration process (for example, after a handling error).

Manual Calibration

Manual calibration is the alternative to the wizard and is documented separately for each station type. Throughout the procedure, record the data you obtain in the calibration register at the end of this guide. The flow is the same for both station families: shared preparation, then the station-specific procedure (Pro or Watch), then shared finalization.

Summary of Steps

Step Watch™ Series Pro Station™ Series
Stop measurement. Preparation, step 1
Record in the calibration register: calibrator's model and serial number, sound level meter's serial number, nominal sensitivity written on the microphone, and calibrator's reference level. See the register, and preparation, step 2
Install and start the calibrator. Preparation, steps 3–5
Note the microphone sensitivity registered in the software. Watch™, step 2 Pro Station™, steps 1–3
Check the measured level and record it in the logbook.* Compare the measured level to the calibrator's reference level.* Watch™, step 3 Pro Station™, step 4
Restore nominal microphone sensitivity. Watch™, step 4 Pro Station™, step 5
Remove the calibrator; note the measured level in the register and compare it to the calibrator's reference level. Watch™, step 7 Pro Station™, step 6
Calibrate the microphone. Watch™, step 7 Pro Station™, step 6
Note the new sensitivity in the calibration register. See the register
Finalize the calibration and resume measurement. Finalization procedure

* These steps are only required during periodic calibration.

Manual Calibration - Preparation

These steps are common to both station types. Complete them first, then continue with the procedure for your station.

  1. Using the monitoring software, stop measurements.

    Watch™ Series

    • Connect a Windows computer or tablet to the station's USB port.
    • Open the Watch™ Utility software and click the Start/Stop button.

    Pro Station™ Series

    • Connect the touchscreen to one of the station's USB ports.
    • Use the Mezzo Noise Monitor software (it should already be open on the station's computer) and press the Stop button.
  2. Verify that the date of the last laboratory calibration — located at the base of the microphone (see Figure B) — is less than one year old. Perform the same check for the reference source.

  3. Remove the windscreen (foam ball) at the microphone base by carefully unscrewing it counter-clockwise. Be careful not to squeeze the windscreen during removal.

    Important. Soft dB cannot be held responsible for damage caused by improper handling of instruments during calibration.

    Windscreen (foam ball) being unscrewed counter-clockwise from the microphone base.
  4. Install the reference source at the end of the microphone until the microphone locks in place. Push the microphone into the receiver base of the reference source and rotate slightly clockwise.

    Reference source being installed onto the end of the microphone.
  5. Wait 90 seconds after installing the reference source, then start the device by pressing the On/Off button (this may vary by model; see the example in the red box).

    Now continue with the procedure for your station: Pro Station™ Series: Step 6 or Watch™ Series: Step 7. Then complete the shared finalization.

    Reference source with its Start button highlighted in a red box.

Manual Calibration - Pro Station™ Series

The following steps make up Step 6 of the manual procedure. Perform them for Pro Station™ Series stations only.

  1. With the touchscreen connected to the station, open the Mezzo Noise Monitor software on the measuring station and click the Stop button to stop the measurement.

    Select the Setup tab. In the Setup tab, click on the Sound Input button or OPL Input button to open the interface for the type of microphone installed.

    Mezzo Noise Monitor with the measurement stopped and the Setup tab highlighted.
  2. Check that the sound level generated by the reference source is lower than the saturation (overload) level of the acoustic sensor displayed in the Peak Overload field (Regular microphone) and Dynamic Range field (SAS microphone).

    SLM Input Setup for a regular microphone; Peak Overload highlighted
    Regular microphone
    SLM Input Setup for a SAS microphone; Dynamic Range highlighted
    SAS microphone
  3. Confirm the microphone information displayed in the interface.

    SLM Input Setup for a regular microphone; Sensitivity and Calibrate highlighted
    Regular microphone
    SLM Input Setup for a SAS microphone; Sensitivity and Calibrate highlighted
    SAS microphone

    Verify that the microphone's serial number, manufacturer name, and model number are correct. This information is on the label at the base of the microphone.

    Note the sensitivity value displayed in the sensitivity field in the calibration register (before calibration) and enter it in the Sensitivity — Before column.

    Click the Calibrate button to open the calibration and level-check interface.

  4. Calibration and level-check interface with the Calibrator Level, Calibrator Frequency, Check, Run, and Measured Level controls.
    1. Depending on the type of reference source, the Calibrator level may need to be adjusted (see the details icon Details icon). It is strongly recommended to verify the reference level against the technical specifications of the reference source and the microphone.
    2. Verify that the reference source is active by confirming a peak at the source's nominal frequency (1000 Hz).
    3. Confirm that Calibrator Level and Calibrator Frequency (point 1) match the reference data written on the calibrator (see Figure C).
    4. Click Check and note the Measured Level value in the calibration register. Click Stop after 5 seconds or more.
    5. Check the difference between the Calibrator Level and the Measured Level:
      1. If the deviation is less than ±1.0 dB, continue the procedure.
      2. If the deviation is greater than or equal to ±1.0 dB, check that the calibrator and microphone are working properly (cables, connectors, and so on). If no apparent problem is found, contact Soft dB. The acoustic sensor and/or calibrator may need to be returned for inspection, potential repair, and/or recertification, if applicable.
  5. Calibration interface
    Sound input setup interface
    1. Return to the Sound Input Setup page by clicking the OK button.
    2. Click the Load From Mezzo button to reset to nominal sensitivity, and note it in the calibration register.
    3. Click the Calibrate button to open the calibration and level-check interface.
  6. Level check:

    Calibration and level-check interface during the level-check step.
    1. Make sure the calibrator is running.
    2. Click the Check button to check the calibration (point 4).
    3. After 5 seconds or more, press the Stop button (point 3).
    4. Check the value displayed in the Measured Level field and enter it in the calibration register.
    5. Repeat the checks presented in step 6.4's sub-step 5

    If the data is within tolerances, proceed to the next step.

  7. Calibrate:

    Calibration and level-check interface during the calibration (Run) step.
    1. Make sure the calibrator is running.
    2. Click the Run button to start calibration (point 2).
    3. After 5 seconds or more, press the Stop button (point 3).
    4. Check the value displayed in the Measured Level field and note it in the calibration register.
    5. Repeat the checks presented in step 6.4's sub-step 5

    If the data is within tolerances, click OK to save this calibration.

  8. Continue to the finalization steps.

Manual Calibration - Watch™ Series

The following steps make up Step 7 of the manual procedure. Perform them for Watch™ Series stations only.

  1. Using a Windows computer or tablet, open the Watch™ Monitoring Utility software and connect the computer to the measurement station.

    Click the Sensor Setup button in the SLM - Microphone column of the Watch™ Utility screen to open the Sensor Setup menu.

  2. Enter the microphone sensitivity value displayed in the Sensitivity field in the calibration register.

    Sensor Setup panel with the microphone sensitivity value highlighted in a green box.
  3. Check that the Dynamic Range value displayed in the Dynamic Value field is higher than the reference source level. Click the Calibrate button to open the calibration interface.

    Sensor Setup panel showing the Dynamic Range value and the control that opens the calibration interface.
  4. As-found level check and restore factory calibration:

    1. Verify that the Level and Frequency parameters are set to the levels indicated on the calibrator, and that the calibrator is enabled.
    2. Start the calibrator.
    3. Click the Check button; a peak at 1000 Hz should be observed.
    4. After more than 5 seconds, note the measured sound level (orange box).
    5. Record the measured sound level in the calibration register.
    6. Check the deviation between the Calibrator Level and the measured level:
      1. If the deviation is less than ±1.0 dB, continue the procedure.
      2. If the deviation is greater than ±1.0 dB, check that the calibrator and microphone are working properly (cables, connectors, and so on). If no apparent problem is found, contact Soft dB. The acoustic sensor and/or calibrator may need to be returned for inspection, potential repair, and recertification, if applicable.
    7. Click the Restore Factory Calibration button.
  5. Post-restore level check:

    1. Start the calibrator.
    2. Click the Check button.
    3. After more than 5 seconds, note the measured sound level (orange box).
    4. Record the measured sound level in the calibration register.
    5. Check the deviation between the Calibrator Level and the measured level:
      1. If the deviation is less than ±1.0 dB, continue the procedure.
      2. If the deviation is greater than ±1.0 dB, check that the calibrator and microphone are working properly (cables, connectors, and so on). If no apparent problem is found, contact Soft dB. The acoustic sensor and/or calibrator may need to be returned for inspection, potential repair, and recertification, if applicable.
  6. Calibrate:

    1. Start the calibrator.
    2. Click the Cali button then wait 5 seconds for the calibration to complete.
    3. After more than 5 seconds, record the measured sound level (orange box).
    4. Record the measured sound level in the calibration register.
    5. Check the deviation between the Calibrator Level and the measured level:
      1. If the deviation is less than ±1.0 dB, continue the procedure.
      2. If the deviation is greater than ±1.0 dB, check that the calibrator and microphone are working properly (cables, connectors, and so on). If no apparent problem is found, contact Soft dB. The acoustic sensor and/or calibrator may need to be returned for inspection, potential repair, and recertification, if applicable.
    6. Click the OK button to save the sensitivity. Do not close with the X or Cancel button, otherwise the updated sensitivity will not be saved.
  7. Continue to finalization steps.

Manual Calibration - Finalization

Once the fields in the calibration register are filled in, finalize the calibration of the acoustic sensor.

  1. Remove the reference source (calibrator) by rotating it slightly clockwise.

    Important. Be careful not to unscrew the microphone membrane.

    Reference source being removed from the microphone by rotating it clockwise.
  2. Reinstall the windscreen (foam ball) by rotating it clockwise onto the membrane.

    Important. Make sure the windscreen is dry and fully covers the microphone.

    Windscreen (foam ball) being reinstalled onto the microphone.
  3. Using the monitoring software, restart measurements.

    Watch™ Series

    • Connect a Windows computer or tablet to the station's USB port.
    • Open the Watch™ Utility software and click the Start button.

    Pro Station™ Series

    • Connect the touchscreen to one of the station's USB ports.
    • Use the Mezzo Noise Monitor software (it should already be open on the station's computer) and click the Run button.

Calibration Register

The calibration register is your record of each calibration: the equipment used, the levels measured before and after, and the resulting sensitivities. Complete it during every calibration — especially for manual-mode calibrations — as described in Recommendations & best practices.

Calibration register — the form to complete during each calibration.

Register legend — colour-coded fields (for example, the green box for the sensitivity before calibration and the orange box for the measured level) correspond to the highlighted areas referenced in the procedures above.